ZeroHour

CVE-2018-8868

CVSS 3.1
6.2 medium
EPSS
<1%p26
Published
()
Modified
Description

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

Vendors
medtronic
Products
24950 mycarelink monitor firmware, 24952 mycarelink monitor firmware
Weakness
CWE-749
Vector
CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L