ZeroHour

CVE-2019-6540

CVSS 3.1
6.5 medium
EPSS
<1%p9
Published
()
Modified
Description

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.

Vendors
medtronic
Products
mycarelink monitor 24950 firmware, mycarelink monitor 24952 firmware, carelink monitor 2490c firmware, carelink 2090 firmware, amplia crt-d firmware, claria crt-d firmware, compia crt-d firmware, concerto crt-d firmware, concerto ii crt-d firmware, consulta crt-d firmware, evera icd firmware, maximo ii crt-d firmware
Weakness
CWE-319
Vector
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N