60
CVE-2019-6540
—CVSS 3.1
6.5 medium
EPSS
<1%p9
Published
()
Modified
Description
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.
- Vendors
- medtronic
- Products
- mycarelink monitor 24950 firmware, mycarelink monitor 24952 firmware, carelink monitor 2490c firmware, carelink 2090 firmware, amplia crt-d firmware, claria crt-d firmware, compia crt-d firmware, concerto crt-d firmware, concerto ii crt-d firmware, consulta crt-d firmware, evera icd firmware, maximo ii crt-d firmware
- Weakness
- CWE-319
- Vector
- CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N